Low cost laserless FTIR spectrometer on the farm for real-time nitrous oxide soil emission measurements

Sammendrag

A low-cost Fourier transform infrared (FTIR) instrument was developed where the traditional He–Ne reference laser was replaced by a low-cost linear encoder. An RMS sampling error of less than 20 nm was achieved by oversampling both the interferogram and the encoder signal and then resampling the interferogram using a correction table for the encoder. A gas calibration model was developed for the system, which was chosen to have a stroke length of 21 mm and, thereby, a resolution of 0.4  cm −1 after apodization. The instrument was mounted on a vehicle and employed in an agricultural field test for measuring soil emissions, in particular nitrous oxide (N 2 O ). The concentration of N 2 O was measured with a root mean squared error of 6 ppb. The results compared well with lab-based gas chromatography measurements.

Kategori

Vitenskapelig artikkel

Språk

Engelsk

Forfatter(e)

Institusjon(er)

  • SINTEF Digital / Smart Sensors and Microsystems
  • Norsk institutt for bioøkonomi

År

2013

Publisert i

Applied Optics

ISSN

0003-6935

Forlag

Optical Society of America

Årgang

52

Hefte nr.

4

Side(r)

93 - 101