Til hovedinnhold
Norsk English

VALIDATION OF A NON-RIGID REGISTRATION METHOD FOR MOTION COMPENSATION IN 4D ULTRASOUND OF THE LIVER

Sammendrag

Future therapy using focused ultrasound (FUS) to treat tumors in abdominal organs, such as the liver, must incorporate motion tracking of these organs due to breathing and drift caused by gravity and intestines (peristalsis). Motion tracking of the target (e.g. tumor) is needed to ensure accurately located sonications. We have performed a quantitative validation of a methodology for motion tracking of the liver with 4D (3D+time) ultrasound. The offline analysis was done using a recently published non-rigid registration algorithm that was specifically designed for motion estimation from dynamic imaging data. The method registers the entire 4D sequence in a group-wise optimization fashion, thus avoiding a bias towards a specifically chosen reference time point. Both spatial and temporal smoothness of the transformations are enforced by using a 4D free-form B-spline deformation model. For our evaluation, three healthy volunteers were scanned over several breath cycles from three different positions and angles on the abdomen (totally nine 4D scans). A skilled physician performed the scanning and manually annotated well-defined anatomic landmarks for assessment of the automatic algorithm. Four engineers each annotated these points in all time frames, the mean of which was taken as a gold standard. The error of the automatic motion estimation method was compared with inter-observer variability. The registration method estimated liver motion better than the observers and had an error (75% percentile over all datasets) of 1 mm. We conclude that the methodology was able to accurately track the motion of the liver in the 4D ultrasound data.

Kategori

Vitenskapelig artikkel

Språk

Engelsk

Forfatter(e)

Institusjon(er)

  • Norges teknisk-naturvitenskapelige universitet
  • Erasmus MC: Universitair Medisch Centrum Rotterdam
  • SINTEF Digital / Helse
  • St. Olavs Hospital HF

År

2013

Publisert i

IEEE International Symposium on Biomedical Imaging

ISSN

1945-7928

Side(r)

792 - 795

Vis denne publikasjonen hos Cristin