Til hovedinnhold
Norsk English

Registration of Real-Time 3-D Ultrasound to Tomographic Images of the Abdominal Aorta

Sammendrag

The purpose of this study was to develop an image-based method for registration of real-time 3-D ultrasound to computed tomography (CT) of the abdominal aorta, targeting future use in ultrasound-guided endovascular intervention. We proposed a method in which a surface model of the aortic wall was segmented from CT, and the approximate initial location of this model relative to the ultrasound volume was manually indicated. The model was iteratively transformed to automatically optimize correspondence to the ultrasound data. Feasibility was studied using data from a silicon phantom and in vivo data from a volunteer with previously acquired CT. Through visual evaluation, the ultrasound and CT data were seen to correspond well after registration. Both aortic lumen and branching arteries were well aligned. The processing was done offline, and the registration took approximately 0.2 s per ultrasound volume. The results encourage further patient studies to investigate accuracy, robustness and clinical value of the approach.

Kategori

Vitenskapelig oversiktsartikkel/review

Språk

Engelsk

Forfatter(e)

Institusjon(er)

  • Norges teknisk-naturvitenskapelige universitet
  • St. Olavs Hospital HF
  • SINTEF Digital / Helse

År

2016

Publisert i

Ultrasound in Medicine and Biology

ISSN

0301-5629

Forlag

Elsevier

Årgang

42

Hefte nr.

8

Side(r)

2026 - 2032

Vis denne publikasjonen hos Cristin